Description
Instructions for use of “Chlorine dioxide” test strips for rapid monitoring of the concentration of working solutions of chlorine dioxide-based disinfectants
1. Purpose
The “Chlorine dioxide” test strips (hereinafter referred to as test strips) are intended for visual monitoring of the preparation and proper storage of working solutions of chlorine dioxide-based disinfectants with concentrations of 0.005; 0.01; 0.02; 0.03; 0.05; 0.1; 0.2; 0.3 wt % ClO2.
The delivery set includes: test strips, color scale, instructions for use, packaging bag.
The test strips are intended for use by personnel responsible for carrying out disinfection measures.
2. Application
Pour 50–100 mL of a well-mixed working solution of the disinfectant at room temperature into a graduated cylinder or a glass beaker. Take a test strip from the package and immerse it for 1 second in the solution so that the indicator zone is completely wetted. Remove the strip from the solution and quickly remove the excess liquid by drawing the edge of the strip along the rim of the glass. Place the strip on white filter paper or a paper napkin with the indicator zone facing up and hold for 30 seconds, after which, within 10 seconds, compare the color of the indicator zone with the color scale to determine the concentration of the working solution.
Table for diluting solutions whose concentrations are not shown on the color scale
Notes:
1. The specified exposure time of the test strips in the analyzed solution and on the filter paper must be observed.
2. Comparison of the color of the test strip with the color scale should be carried out under normal workplace lighting conditions. The appearance of halos on the test strips that are not shown on the color scale should be considered insignificant.
3. The determination of the concentration of the working solution should be performed three times. The result is considered reliable if it is the same in at least two repeated determinations. If the analysis needs to be repeated, use a fresh portion of the solution.
Caution:
– Vapors of chemicals may affect the determination results; therefore, avoid chemical vapors during storage and use of the test strips;
– Remove from the pack only the number of test strips needed for the determination; do not leave the pack open;
– Do not remove excess solution from the test strips with filter paper;
– Use the enclosed color scale only for this package of test strips;
– Do not expose the color scale to direct sunlight or disinfectants;
– Do not use any portion of the solution used for concentration testing for disinfection purposes.
3. Safety requirements
The test strips do not release toxic substances into the environment and do not have a harmful effect on the human body upon direct contact. No special safety or environmental protection measures are required when working with the test strips. After use, the test strips are to be disposed of as household waste.
4. Storage and transportation
The test strips are stored and transported in the manufacturer’s packaging at temperatures from -15°C to +30°C and humidity not exceeding 80%, without exposure to chemical vapors. Shelf life in the unopened manufacturer’s packaging is 18 months. The shelf life of the test strips after the initial opening of the package is 3 months.